The Annex to Decision 2002/364/EC is amended in accordance with the Annex to this Decision.
資料由法律人 LawPlayer整理提供·EU law / curated by LawPlayer from EUR-Lex
Commission Implementing Decision (EU) 2020/350 of 28 February 2020 amending Decision 2002/364/EC as regards definitions of first–line assays and confirmatory assays, requirements for devices for self-testing and requirements for HIV and HCV rapid tests, confirmatory and supplementary assays (notified under document C(2020) 1086) (Text with EEA relevance)
1. This Decision shall apply from 2 March 2021.
2. Notwithstanding paragraph 1, from 2 March 2020 until 1 July 2020 Member States shall apply the presumption of compliance laid down in Article 5(3) of Directive 98/79/EC for all in vitro diagnostic medical devices that comply with any of the following:
(a)
the common technical specifications laid down in Decision 2002/364/EC as amended by Commission Decision 2011/869/EU ( 4 ) ;
(b)
the common technical specifications laid down in Decision 2002/364/EC as amended by Commission Implementing Decision (EU) 2019/1244 ( 5 ) ;
(c)
the common technical specifications laid down in Decision 2002/364/EC as amended by this Decision.
3. Notwithstanding paragraph 1, from 2 July 2020 until 1 March 2021 Member States shall apply the presumption of compliance laid down in Article 5(3) of Directive 98/79/EC for all in vitro diagnostic medical devices that comply with either of the following:
(a)
the common technical specifications laid down in Decision 2002/364/EC as amended by Implementing Decision (EU) 2019/1244;
(b)
the common technical specifications laid down in Decision 2002/364/EC as amended by this Decision.
This Decision is addressed to the Member States.
Schedules & Appendices
ANNEX
The Annex to Decision 2002/364/EC is amended as follows:
1.
Section 2 is amended as follows:
(a)
the following definition of ‘First line assay’ is inserted between the definition of ‘Whole system failure rate’ and the definition of ‘Confirmation assay’:
‘
First-line assay
First-line assay means an assay used to detect a marker or analyte, and which may be followed by a confirmatory assay. Devices intended solely to be used to monitor a previously determined marker or analyte are not considered first-line assays.’;
(b)
the definition of ‘Confirmation assay’ is replaced by the following:
‘
Confirmatory assay
Confirmatory assay means an assay used for the confirmation of a reactive result from a first-line assay.’;
2.
Section 3 is amended as follows:
(a)
Sub-section 3.1.1 is replaced by the following:
‘3.1.1.
Devices which detect virus infections shall meet the requirements for sensitivity and specificity set out in Table 1, Table 3, Table 4 and Table 5, which apply to them taking account of the intended purpose of the devices concerned, virus type and entities to be detected (antigen and/or antibody). See also principle 3.1.11 for first-line assays.’;
(b)
Sub-section 3.1.3 is replaced by the following:
‘3.1.3.
Devices for self-testing shall meet the same CTS requirements for sensitivity and specificity as respective devices for professional use. Relevant parts of the performance evaluation shall be carried out (or repeated) by appropriate lay users to validate the operation of the device and the instructions for use. The lay users selected for the performance evaluation shall be representative of the intended users groups.
Performance evaluation of a device for self-testing shall include, for each body fluid claimed for use with the device, e.g. whole blood, urine, saliva, etc., at least 200 lay users that are known positive for the infection and at least 400 lay users that do not know their status, of which at least 200 are at high risk of acquiring the infection. The sensitivity and specificity of the device for self-testing in the hands of lay users shall be defined against the confirmed patient infectious status.’;
(c)
Sub-section 3.1.9 is replaced by the following:
‘3.1.9.
Performance evaluation of first line assays shall include 25 positive (if available in the case of rare infections) ‘same day’ fresh serum samples (≤ 1 day after sampling).’;
(d)
Sub-section 3.1.11 is replaced by the following:
‘3.1.11.
For performance evaluations for first line assays (Table 1 and Table 3) blood donor populations shall be investigated from at least two blood donation centers and consist of consecutive blood donations, which have not been selected to exclude first time donors.’;
(e)
Sub-section 3.4.2 is replaced by the following:
‘3.4.2.
The manufacturer’s batch release testing for first line assays shall include at least 100 specimens negative for the relevant analyte.’.
3.
Table 1 is replaced by the following:
‘Table 1
First-line assays, excluding rapid tests: anti-HIV 1/2, HIV 1/2 Ag/Ab, anti-HTLV I/II, anti-HCV, HCV Ag/Ab, HBsAg, anti-HBc
anti-HIV 1/2, HIV 1/2 Ag/Ab
Anti-HTLV-I/II
anti-HCV, HCV Ag/Ab
HBsAg
Anti-HBc
Diagnostic sensitivity
Positive specimens
400 HIV-1
100 HIV-2
including 40 non-B-subtypes, all available HIV/1 subtypes shall be represented by at least 3 samples per subtype
300 HTLV-I
100 HTLV-II
400 (positive samples)
Including samples from different stages of infection and reflecting different antibody patterns.
Genotype 1-4: > 20 samples per genotype (including non-a subtypes of genotype 4);
5: > 5 samples;
6: if available
400
including subtype-consideration
400
including evaluation of other HBV-markers
Sero-conversion panels
20 panels
10 further panels (at notified body or manufacturer)
To be defined when available
20 panels
10 further panels (at notified body or manufacturer)
20 panels
10 further panels (at notified body or manufacturer)
To be defined when available
Analytical sensitivity
Standards
0,130 IU/ml ( WHO International Standard: Third International Standard for HBsAg, subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226)
Specificity
Unselected donors (including first-time donors)
5 000
5 000
5 000
5 000
5 000
Hospitalized patients
200
200
200
200
200
Potentially cross-reacting blood-specimens (RF+, related viruses, pregnant women, etc)
100
100
100
100
100’
4.
Table 3 is replaced by the following:
‘Table 3
Rapid tests: anti-HIV 1/2, HIV 1/2 Ag/Ab, anti-HCV, HCV Ag/Ab, HBsAg, anti-HBc, anti-HTLV I and II
anti-HIV 1/2, HIV 1/2 Ag/Ab
anti-HCV, HCV Ag/Ab
HBsAg
anti-HBc
anti-HTLV I and II
Acceptance criteria
Diagnostic sensitivity
Positive specimens
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Sero-conversion panels
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Same criteria as in Table 1
Diagnostic specificity
Negative specimens
2 1 000 blood donations
1 000 blood donations
1 000 blood donations
1 000 blood donations
1 000 blood donations
≥ 99 %
(anti-HBc: ≥ 96 %)’
200 clinical specimens
200 clinical specimens
200 clinical specimens
200 clinical specimens
200 clinical specimens
200 samples from pregnant women
200 samples from pregnant women
200 samples from pregnant women
200 samples from pregnant women
100 potentially interfering samples
100 potentially interfering samples
100 potentially interfering samples
100 potentially interfering samples
100 potentially interfering samples
5.
Table 4 is replaced by the following:
‘Table 4
Confirmatory and supplementary assays for anti-HIV 1/ 2, HIV 1/2 Ag/Ab, anti-HTLV I and II, anti-HCV, HCV Ag/Ab, HBsAg
anti-HIV 1/2, HIV 1/2 Ag/Ab confirmatory assays
anti-HTLV I and II confirmatory assays
anti-HCV, HCV Ag/Ab supplementary assays
HBsAg confirmatory assays
Acceptance criteria
Diagnostic sensitivity
Positive specimens
200 HIV-1 and 100 HIV-2
Including samples from different stages of infection and reflecting different antibody patterns
200 HTLV-I and 100 HTLV-II
300 HCV (positive samples)
Including samples from different stages of infection and reflecting different antibody patterns.
Genotypes 1 – 4: > 20 samples (including non-a subtypes of genotype 4;
Genotype 5: > 5 samples;
Genotype 6: if available
300 HBsAg
Including samples from different stages of infection
20 ‘high pos’ samples (>26 IU/ml); 20 samples in the cut-off range
Correct identification as positive (or indeterminate), not negative
Sero-conversion panels
15 sero-conversion panels/low titre panels
15 sero-conversion panels/low titre panels
15 sero-conversion panels/low titre panels
Analytical sensitivity
Standards
Third International Standard for HBsAg, subtypes ayw1/adw2, HBV genotype B4, NIBSC code: 12/226
Diagnostic specificity
Negative specimens
200 blood donations
200 blood donations
200 blood donations
10 false positives as available from the performance evaluation of the first line assay ( 1 ) .
No false-positive results/ ( 1 ) no neutralisation
200 clinical samples including pregnant women
200 clinical samples including pregnant women
200 clinical samples including pregnant women
50 potentially interfering samples, including samples with indeterminate results in other confirmatory assays
50 potentially interfering samples, including samples with indeterminate results in other confirmatory assays
50 potentially interfering samples, including samples with indeterminate results in other supplementary assays
50 potentially interfering samples
( 1 ) Acceptance criteria: no neutralisation for HBsAg confirmatory assay.’
Cite this act
Commission Implementing Decision (EU) 2020/350 of 28 February 2020 amending Decision 2002/364/EC as regards definitions of first–line assays and confirmatory assays, requirements for devices for self-testing and requirements for HIV and HCV rapid tests, confirmatory and supplementary assays (notified under document C(2020) 1086) (Text with EEA relevance) (EUR-Lex). Retrieved via LawPlayer, https://lawplayer.com/eu/act/32020D0350
© European Union, https://eur-lex.europa.eu, 1998-2026. Reuse authorised under Commission Decision 2011/833/EU, provided the source is acknowledged.
本頁資料來源:EUR-Lex·整理提供:法律人 LawPlayer· lawplayer.com